WISE Study Unruptured aneurysm subgroup:

WEB™ Implant Safety and Long Term Effectiveness | 1-year unruptured aneurysm results
WEB™ Intrasaccular Embolization

WISE study Unruptured aneurysm sub group demonstrates "The WEB™ device is safe and effective for the treatment of unruptured intracranial aneurysms with high rates of adequate occlusion, low rates of recurrence and retreatment, and reliable protection from rupture."

Patients

458

Countries

US, Germany, France, Finland, UK, Denmark, Belgium, Hungary, Turkey, Canada, China

Clinical Centers

58

Design

Pooled analysis of individual subject-level data from 6 prospective, multicenter, core laboratory adjudicated, externally monitored studies evaluating the WEB device for the treatment of unruptured WNBAs.

Scope

Patients with unruptured wide neck bifurcation intracranial aneurysms

Purpose

This study reports data from the largest prospectively collected, fully adjudicated data set of WEB™ device-treated subjects and further confirms the performance of WEB™ device for WNBA treatment. These high-quality data are useful in confirming or refuting multiple factors thought to impact rates of occlusion, retreatment and procedural thromboembolic events.

Efficacy results

Technical success: 98.0%

At 12 months:

  • Complete occlusion: 54.7%

  • Adequate occlusion: 82.8%

  • Retreatment: 2.5%

  • Recurrence: 6.4%

  • Parent artery stenosis>50%: 0.8%

Sub group analysis performed by aneurysm size (≤6 mm, >6 mm):

  • Complete occlusion: 67.5% Aneurysm size ≤6 mm

  • Adequate occlusion: 88.1% Aneurysm size ≤6 mm

Safety results

Procedural TE: 4.2% (symptomatic 1.3%)

At 1 month:

  • New Bleed: 0%

  • Stroke: 5.2% (disabling 0.7%)

  • Mortality: 0% (treatment-related 0%)

  • Morbidity: 0.9% (treatment-related 0.4%)

At 12 months:

  • New Bleed: 0 %

  • Stroke: 6.6% (disabling 0.9%)

  • Mortality: 1.3% (treatment-related 0.2%)

  • Morbidity: 1.2% (treatment-related 0.5%)

  • Mortality WEB related: 0%

  • Morbidity WEB related: 0%

Please refer to Instruction for Use (IFU) in the specific market/country that you are looking into:

MM2467 WW 10/26

Explore more studies

CLEVER Study: CLinical EValuation of WEB™ 0.017 Device in Intracranial AneuRysms

WEB™ Intrasaccular Embolization

WEBCAST 2 Study: Extension of WEBCAST study

WEB™ Intrasaccular Embolization

WEBCAST Study: Clinical Assessment of IntraSaccular Aneurysm Therapy

WEB™ Intrasaccular Embolization

Request more information

The information you provide through this form is necessary to respond to your request and may be shared with the relevant Terumo Neuro group entity or entities responsible for handling it. For more information on how we process your personal data and on your rights, please refer to our privacy notice.