CASPER™ Indications, Safety, and Warnings

See Instructions for Use for a complete list of warnings, precautions, and contraindications

MRI Safety Information

MR Conditional 

Non-clinical testing1 demonstrated that the Carotid Stent is MR Conditional. A patient with this device can be scanned safely in an MR system immediately after placement under the following conditions:

- Static magnetic field of 1.5-Tesla and 3-Tesla, only

- Maximum spatial gradient magnetic field of 720-Gauss/cm (72-T/m)

- Maximum MR system reported, whole body averaged specific absorption rate (SAR) of 2-W/kg for 15 minutes of scanning (i.e., per pulse sequence) in the Normal Operating Mode of operation for the MR system

- Under the scan conditions defined, the Carotid Stent is expected to produce a maximum temperature rise of 1.7°C after 15-minutes of continuous scanning (i.e., per pulse sequence).

Artifact Information

In non-clinical testing, the image artifact caused by the Carotid Stent extends approximately 5-mm from this implant when imaged using a gradient echo pulse sequence and a 3-Tesla MR system.

1Test Report TR12-203 on file at Terumo Neuro.